Alerts;
This is not a comprehensive list of warnings and alerts for this system. It is just a overview of some that have been found to impact worker and patient safety in the past. When in doubt about the meaning or importance of a warning or alert, refer to the operator's manual or contact your service agent.
Note that a repeating tone during x-ray usually indicates activation of High Level Fluoroscopy (HLF). Only use HLF when other acquisition or fluoroscopy modes are insufficient to penetrate the patient.
This is not a comprehensive list of warnings and alerts for this system. It is just a overview of some that have been found to impact worker and patient safety in the past. When in doubt about the meaning or importance of a warning or alert, refer to the operator's manual or contact your service agent.
Note that a repeating tone during x-ray usually indicates activation of High Level Fluoroscopy (HLF). Only use HLF when other acquisition or fluoroscopy modes are insufficient to penetrate the patient.
4 HOUR WARM-UP REQUIRED - PRESS ANY KEY:
If the system has been stored for more than 60 days, a warm-up period is necessary for ion removal from the image intensifier tube. Leave the system ON with no operation for 4 hours. Press any C-Arm control panel key to continue.
|
24 HOUR RECHARGE REQUIRED - X-RAYS DISABLED:
If the system has been in storage for more than 6 months, the batteries need recharging. X-rays are disabled. Leave the WorkStation power cord plugged in, make sure the interconnect cable is properly connected, and wait 24 hours before use. The system does not need to be turned on to recharge.
|
AD CHANNEL # FAIL:
The analog-to-digital channels are checked during system start-up. If a fault in the data conversion process is detected in any of the channels, the system will not operate. Perform the error recovery steps. If the problem persists, call a qualified service person.
|
ANODE IS HOT - XX%:
This message alternates with the technique displayed on the control panel display and an alarm sounds. The anode temperature is at 80% of its rated heat capacity or greater. The alarm may be disabled by pressing the ALARM RESET switch on the control panel. Continued use without cooling may damage the x-ray tube. Although fluoroscopy is not prevented, you should wait for the tube to cool before making another exposure. HLF and film exposures are not allowed. WARNING: When the ANODE IS HOT - XX% message displays, personnel should avoid bodily contact with the X-ray tube. Do not allow the X-ray tube housing to contact the patient.
|
ANODE IS WARM - XX%:
This message will alternate with the technique displayed on the control panel display. Anode temperature is at 70% of its rated heat capacity or greater. You may continue with fluoroscopy, but discretion is advised. WARNING: When the ANODE IS WARM -XX% message displays personnel should prevent all bodily contact with the X-ray tube.
|
ARMED:
This message is displayed as the system prepares the technique for a film exposure. If the X-ray switch has not been pressed after 60 seconds the message will be removed from the display.
|
BATTERY CHARGE - XX%:
If the effective battery charge drops to between 70 to 40 percent, this message will alternate with the technique displayed on the control panel display. FLUORO, FIL and HLF exposures are still allowed. To charge the batteries, leave the system plugged into an AC receptacle with the interconnect cable in place between the C-Arm and the WorkStation. The WorkStation power switch should be turned to the OFF position. Normally, a full battery recharge requires only a few hours. If the batteries are further discharged without adequate recharge time, the effective charge may drop below 70 percent.
|
BATTERY CHARGE - XX% /WAIT:
If the effective battery charge drops below 40 percent, this message will alternate with the technique displayed on the control panel display. The WAIT message appears in the center of the technique display. Exposures are allowed in FLUORO mode but FILM and FLF exposures are not allowed until the battery has recharged. To charge the batteries, leave the system plugged into an AC receptacle with the interconnect cable in place between the C-Arm and the WorkStation. The WorkStation power switch should be turned to the OFF position.
|
CHARGER FAILED:
If the battery charger fails at boot-up this error is displayed and the system will not operate. Perform the error recovery steps described at the beginning of this chapter. If the failure occurs during the operation the message will alternate with the technique displayed on the control panel. Continued use will result in the system failure. Call your service representative as soon as possible.
|
COLLIMATOR CAL REQUIRED - PRESS ANY KEY:
The software has determined that the current collimator iris or leave position values do not match the reference values stored in memory. Press any key to continue. Call your service representative as soon as possible.
|
COL IRIS POTENTIOMETER ERROR - PRESS ANY KEY:
The software has sensed that the collimator iris potentiometer is not working and therefore the iris position cannot be determined. Press any key to continue. Call your service representative as soon as possible.
|
COL IRIS TOO LARGE:
The collimator iris is larger than the control panel indicates. This message will alternate with the technique display. CAUTION: Continuing system operation when the COL IRIS TOO LARGE message is displayed may result in over exposing the patient due to a larger than indicated field size. Complete the current procedure if necessary, then call for service.
|
COL IRIS UNSTABLE:
The iris collimator motor control cannot maintain position tolerance. This message alternates with the technique displayed on the control panel display. You may continue to use the system, although the iris instability may degrade the image aat the edges of the X-ray field. Complete the current procedure if necessaary and then call your service representative.
|
COLLIMATOR STUCK:
This message alternates with the technique displayed on the control panel display. The collimator iris motor is unable to open or close the collimator iris. You may continue to use the system, although you will not be able to adjust the collimator iris. Complete the current procedure if necessary and then call your service representative.
|
COMMUNICATION FAILED:
System communication has failed. The Mobile C-Arm is not allowed to take X-rays under these conditions. Wait for approdximately one minute for this condition to clear and if it does not, perform the error recovery steps detailed earlier.
|
CONTROL PANEL ERROR:
Communication to the control panel has been lost, terminating system operation. Perform the error recovery steps described earlier.
|
DATA ERROR:
Corrupted software or data has been detected. X-rays are disabled. Perform the error recovery steps detailed earlier.
|
FAST STOP ACTIVATED:
This message will display after a FAST STOP button has been pressed. You must cycle the WorkStation power switch to OFF and then ON to restart the system. If you did not press a FAST STOP button and this message is displayed a haardware or software fault has occurred. Perform the error recovery steps described previously.
|
FILAMENT CAL REQUIRED - PRESS ANY KEY:
The system has sensed that the filaments have not been calibrated. It is possible to received multiple mA errors if this condition exists. Call your service representative as soon as possible to perform a calibration. WARNING: If the procedure is continued when the FILAMENT CAL REQUIRED message is displayed, the patient may receive a higher dose than that indicated. Press any C-Arm control panel key to continue with the procedure if necessary, then call for service.
|
FILAMENT REGULATOR FAILURE - PRESS ANY KEY:
The filament current has been sensed as out of tolerance. You may press any C-Arm control panel key to continue. However, the resolution of future images may be degraded. Contact your service representative as soon as possible. WARNING: If the procedure is continued when the FILAMENT REG FAILURE message is displayed, the patient may receive a higher dose than that indicated. Press any C-Arm control panel key to continue with the procedure if necessaary, then call for service.
|
FILAMENT SELECT ERROR - PRESS ANY KEY:
The filament size selected by the software does not match the current filament in the PIO hardware. Press any C-arm control panel key to continue. X-rays are disabled while the message is displayed..
|
HLF OVERTIME:
The HLF (High Level Fluoro) or Digital Cine exposure has exceeded the preset time interval and has been terminated. The time interval parameters are dependant on the pulses per second selected. This safety precaution discourages excessive continuous lengths of time in HLF mode. This message remains displayed until the footswitch is released.
|
HOUSING IS HOT - XX%:
This message alternates with the technique displayed on the control panel display and an alarm sounds. The housing temperature is aat 80% of its rated heat capacity or greater. The alarm may be disabled by pressing the ALARM RESET switch on the control panel. Although fluoroscopy is not prevented, you should wait for the tube to cool before making another exposure. HLF and Film exposures are not allowed at this temperature. CAUTION: Continued use without cooling may damage the X-ray tube. Although fluoroscopy is not prevented, you should wait for the tube to cool before making another exposure. Personnel should avoid bodily contact with the X-ray tube housing when this message is displayed. Do not allow the X-ray tube housing to contact the patient.
|
HOUSING OVERHEATED:
The housing temperature has reached 100% of its rated capacity and the thermal cutoff switch has activated (opened). Operation is terminated. Let the housing cool. CAUTION: When the HOUSING OVERHEATED messaage appears, the X-ray tube housing is extremely hot and must be allowed to cool before taking more exposures. Avoid bodily contact with the X-ray tube housing. Do not allow the X-ray tube housing to contact the patient.
|
HOUSING WARM - XX%:
The housing temperature is at 70% of its rated heat capacity or greater. This message alternates with the technique displayed on the control panel display. You may continue with fluoroscopy, but discretion is advised. WARNING: When the HOUSING WARM - XX% messaage appears, personnel should avoid bodily contact with the X-ray tube housing. Do not allow the X-ray tube housing to contact the patient.
|
HV GENERATOR ERROR:
Software has detected an error in the High Voltage Generator. The system automatically shuts down, preventing operation. Perform the error recovery steps described earlier.
|
HV REGISTER FAIL:
Software has detected a failure in the high voltage register. The system automatically shuts down, preventing operation. Perform the error recover steps described earlier.
|
INTERLOCK FAILURE:
The interlock circuit has failed during system start-up. Reboot the system and if the message is displayed again, call for service.
|
KEY STUCK - RELEASE, THEN PRESS ANY KEY TO CONTINUE:
A key press has been sensed on the control panel during boot-up, possibly indicating a stuck control panel key. Verify that there are not any objects pressing against the control panel. After releasing the stuck key, press any C-arm control panel key to continue. A malfunctioning control panel key may interfere with operation if you attempt to continue. Contact your field service representative as soon as possible. .
|
KEYSWITCH IN STANDBY - XRAYS AND MOTORS DISABLED:
This message will display whenever the X-ray keyswitch located on the Mobile C-Arm is turned to the STANDBY position. Turn the keyswitch to the ON position to operate the system. If the keyswitch is already in the ON position there may be a fault in the keyswitch or software. In this case perform the error recovery steps described previously.
|
KV ON IN ERROR:
High voltage is being generated without an X-ray switch being activated, indicataing a fault with the high voltage generator. The system will not operate with this error. Perform the error recovery steps described previously.
|
LIFT SWITCH STUCK:
If the lift switch is pressed continuously for 30 seconds this message will display briefly. This may indicate a foreign object is pressing against the switch or a faulty switch or lift circuit. If there is no object pressing against the switch, perform the error recovery steps described previously. WARNING: In the event of an uncommanded vertical column movement, immediately move the Mobile C-Arm out of the patient environment until the problem has been corrected.
|
MA ON IN ERROR:
X-ray current (tube current) has been detected without an X-ray switch being activated, indicating a fault with the X-ray generator. The system will not operate with this error. Perform the error recovery steps described previously.
|
MA SENSOR FAIL:
The system has detected a failure in one of the mA sensors. This message will alternate with the technique displayed on the control panel display. Perform the error recovery steps described previously.
|
OVERLOAD FAULT:
This may indicate a fault in the high voltage regulator circuit. The first time this fault is detected, the message appears only briefly after an exposure. The second time this fault occurs, the system automatically shuts down and the message remains on the display. Perform the error recovery steps described previously.
|
OVERVOLTAGE FAULT:
This may indicate an X-ray generator failure. The first time this fault is detected, the message appears only briefly after an exposure. The second time this fault occurs, the system automatically shuts down and the message remains on the display. Perform the error recovery steps described previously.
|
PLEASE WAIT:
This message may be displayed briefly on the control panel when the system is updating generator data or performing internal tests. The message should clear within moments. If the message does not clear within 30 seonds to a minute, perform the error recovery steps described previously.
|
PRECHARGE CIRCUIT TIMEOUT:
The CONTACT relay has failed to close during the start-up process. This condition results in automatic system shutdown. Perform the error recovery steps described previously.
|
PRECHARGE VOLTAGE ERROR:
The precharge voltage has been sensed as too high during the start-up process. This condition results in automatic system shutdown. Perform the error recovery steps described previously.
|
RELEASED EARLY:
During a FILM exposure, the X-ray switch was released before the desired mAs was reached. The exposure time may have been too short for a good image. This message appears briefly at the end of the terminated exposure. Using a new film in the cassette, press the X-ray switch and allow enough time for the system to terminate the exposure.
|
SATURATION FAULT:
Indicates a fault in the high voltage regulator circuit. The system will not operate with this error. Perform the error recovery steps described previously.
|
STATOR NOT ON:
The X-ray tube stator is not on, and the tube's anode cannot rotate. The system will not operate with this error. Perform the error recovery steps described previously.
|
WARNING, HIGH KV - PRESS ANY KEY:
An error has been detected in the kV loop. The actual kVp is higher than that indicated on the control panel display and is not within specified tolerances. WARNING: Continuing with the procedure when the WARNING, HIGH KV message has been displayed may subject the patient to a higher dose than that indicated. Press any C-Arm control panel key to continue with the procedure
|
WARNING, HIGH MA - PRESS ANY KEY:
A calibration error has been detected. The actual mA is higher than that indicated on the control panel display and is not within specified tolerances. WARNING: Continuing with the procedure when the WARNING, HIGH MA message has been displayed may subject the patient to a higher dose than that indicated. Press any C-Arm control panel key to continue with the procedure.
|
WARNING, LOW KV - PRESS ANY KEY:
An error has been detected in the kV loop. The actual kVp is lower than that indicated on the control panel display and is not within specified tolerances. Lower kVp may result in poor image quality. Press any C-Arm control panel key to continue.
|
WARNING, LOW MA - PRESS ANY KEY:
A calibration error has been detected. The actual mA is lower than that indicated on the control panel display and is not within specified tolerances. Lower mA may result in poor image quality. Press any C-Arm control panel key to continue.
|
X-RAY OVERTIME - PRESS ANY KEY:
The film shot has continued beyond the time required to achieve the correct exposure and software has terminated the exposure. Press any control panel key to continue. WARNING: When the X-ray overtime message is displayed the patient may have received a higher dose than expected and continued film exposures may result in a higher than expected dose and longer exposure times.
|
X-RAY SWITCH SECURITY ERROR:
A mismatch between the handswitch or footswitch and a security line has been sensed. System operation is terminated. Call your service representative.
|
X-RAY SWITCH STUCK:
One of the X-ray exposure switches has been detected as enabled during the boot sequence. Verify that switches are not being enabled by a foreign object. Or disconnect the handswitch and/or footswitch and attempt restarting the system. If the system restarts, the disconnected device may contain a fault requiring service. If the message persists, there is an internal fault that is preventing system operation. Call your service representative.
|
X-RAY TUBE TEMPERATURE SENSOR FAILURE - PRESS ANY KEY:
The condition of the X-ray tube's heat sensing element is tested during start-up. This message appears if the test indicates that the element is defective. You may continue using the system, but there is a danger that the X-ray tube may overheat during use without further warnings being given. Press any C-arm control panel key to acknowledge the message and call your service representative.
|