Introduction
In most of the modalities discussed in this manual, film is used and the final link in the diagnostic imaging chain, photographic processing, is very important and often the one that is overlooked. All of the efforts by the technical staff aimed at maintaining quality imaging equipment will be futile, if the photographic processor is not maintained under a good quality control program. The photographic process controls not only the image quality (eg. contrast, density, base plus fog) but also the patient exposure.
This chapter deals not only with the photographic processor and all of the associated aspects of film processing including darkroom, fog, daylight loading systems, darkroom conditions, film and chemical storage, but also with daily processor quality control.
Definition
Photographic quality control is a process that ensures the optimization of film processing in order to maintain high image quality and low radiation exposure to patients.
In a photographic quality assurance program the following information should be logged as appropriate:
All of the above information should be included on the sensitometry charts in an effort to assess continuity in contrast and density.
Processing Techniques
Standards
Manual Processors
Sight developing must never be done. The time-temperature method of manual processing must be followed according to the chemical manufacturer's specification. The accuracy of the timer and thermometer should be checked monthly.
Manual developer containers must be opaque and must be covered to prevent deterioration or oxidation of the developer. A non-mercury thermometer and a minute timer are necessary for proper manual processing. A thermometer on the in-coming hot water line (mixing valve), or a thermometer external to the developer tank may not accurately measure the developer temperature.
Renewal of chemistry depends on workload, original volume, replenishment (if any), on the appearance of the radiographs before a need to increase patient exposure is in order to obtain a good film quality.
Quality control checks, ie. sensitometry, should be performed to ensure the activity level of the chemicals before processing clinical films.
Automatic Processing
Automatic processors should be cleaned regularly and the cross-over rollers should be cleaned each operational day. Digital readouts of developer temperature or temperature gauges that are part of the processor itself should not be used as the sole check on developer temperature as they have been seen to be inaccurate. A separate thermometer is required to monitor the developer's temperature.
Dental Processors
Not all processors are fully automatic. Some have automatic transportation systems, but use manual chemicals because they do not have a heater to warm the developer.
Processors designed for use at room temperature do not contain a heating element. 21 As a result, special attention should be given to the location of such processors. A draft from a window may cause a fluctuation in the temperature of the chemicals. Overheated, small enclosures with improper air circulation may make the developer too warm. Both of these situations can result in poor quality images and increase the potential for repeat radiographs or exposures.
21 Some automatic processors without heating elements use manual chemicals.
Evaluate as follows:
Automatic Processors
The developer temperature, processing chemistry, cycle time of the automatic processor, and radiographic exposure, are factors which influence the quality of the final image. In order to obtain high quality films, the manufacturer's recommendations concerning the operation of the processor should be closely followed.
When measuring developer temperature a separate thermometer must be uses as it is not acceptable to rely solely on the integrated digital readouts of developer temperature or of the temperature gauges that are part of the processor, as these can be out of calibration by as much as 5 degrees Celsius.
Equipment
22 A non-mercury thermometer (i.e. alcohol or metal) is recommended, in order to prevent contamination of the processor or chemicals by mercury in the event of thermometer breakage.
Procedure
In addition, it is important that you follow the manufacturer's recommendations for proper maintenance of your processor by:
Manual Processing
The time/temperature method of processing films must be followed, according to the chemical manufacturer's directions. Sight development must NEVER be done as it promotes the use of higher radiation exposure to patients and often results in poor quality radiographs.23
23 Proper photographic chemical tanks (opaque) should be used.
The tanks should have a lid to cover the chemicals to minimize oxidation.
Solutions should be changed at regular intervals as recommended by the manufacturer.
Solutions should be topped up or replenished according to manufacturer's instructions.
Equipment
Procedure
Sensitometry Analysis
Optional with dental equipment.
Sensitometry is performed on a daily basis to ensure that all processors are operating at the same levels and producing consistent, high-quality films.
Equipment
Procedure
Dental Step Wedge Analysis (Sensitometry)
Frequency
Weekly 24
24 Each x-ray machine should yield films of similar density and contrast. However, for photographic quality assurance and machine output reproducibility assessment, only compare the films taken for each x-ray machine.
Equipment
Procedure
Note:
A visible difference in corresponding step images indicates an output difference of approximately +/- 20 percent. It is possible that the patient may be receiving an overexposure.
Safelights
Standard
A safelight must have the correct filter for the film type being used and the bulb must be of the correct wattage. A direct safelight cannot exceed 15 watts. An indirect safelight cannot exceed 25 watts.
Safelights must be placed a minimum of 1 metre from the working area.
The filter must not be cracked or scratched or otherwise permit unfiltered light (white light) to reach the film. The filter must be of a type specific to the film being used.
Evaluation
Dental automatic processors are frequently fitted with a light-tight compartment covered by a filter screen which enables the dentist to use the processor in a location other than a darkroom. The operator slips his/her hands through a set of sleeves and views the unwrapping of the film and the insertion of it into the processor by looking through the filter screen.
Darkroom Integrity
Standard
There must be no white light shining in the darkroom when film are being transferred to and from their protective holders.
Evaluation
Close the door to the darkroom. Turn out the lights, allow your eyes to adjust to the dark and check the darkroom for light leakage. In particular, check around the vertical edge of the door frame, window blinds and pass boxes. On automatic processors, which have daylight loading, check entry sleeves for tears or cracks that may cause film fogging.
Also refer to darkroom testing in mammography section for more information.
Procedure
Evaluate as follows
The exposure should produce less than a 0.05 increase in the mid-density portion of the film. Ideally, less than 0.05 increase should also be obtained with a two minute exposure to the darkroom safelight.
Unused Film Storage
Standard
If unused film is stored in the x-ray room, then it must be stored in a container which provides adequate protection against scatter radiation such as a lead container or something equivalent to 1.6 mm of lead. Film should be stored standing on its edge, not flat, as it is sensitive to pressure marks. Outdated film must not be used as it has an increased base fog.
Photographic materials should be stored at temperatures less than 24 degrees C, preferably in the range of 15 to 21 o C. Open packages of film should be stored in an area with humidity ranging between 40 and 60 per cent.
Many x-ray facilities provide lead shielding on the walls to the seven foot level. Care should be taken to ensure either that film is not stored above this level in darkrooms adjacent to exposure rooms or that the lead shielding is extended to the ceiling in such areas.
Evaluation
If in doubt about the safety of the film storage, take a scatter measurement at the storage position. Any reading will indicate the film is being exposed to scatter radiation. If there is no reading, but there is a possibility that the film could be fogged (tube near or pointed at film), a direction warranted.
Film can be stored in a freezer to extend shelf life. The expiry date on the box must be changed, so that the duration the film remains in the freezer is not included in the shelf life.
Table 12-1 Summary of Processor Maintenance
Procedure |
Frequency |
Comments |
Process sensitometric strip and measure the temperature of the developer daily |
At least once per day and including week-ends if applicable |
Strips should also be run when the processor is cleaned or any maintenance is performed |
Wash crossover assemblies |
At least once per day |
Clean when processor is turned on and turned off |
Check replenishment rates |
Daily |
|
Remove and wash all racks |
Weekly |
Inspect racks for damage |
Check recirculation pumps |
Weekly |
Developer and fixer |
Check replenishment pumps |
Weekly |
Developer and fixer |
Check fixer and wash temperatures |
Weekly |
Use a calibrated thermometer placed in the tank |
Check processing time |
Weekly |
Dry to drop |
Drain and clean tanks |
Monthly |
Remember to use starter |
Check and replace filters |
Monthly |
|
Clean intensifying screens |
Semi-annually |
Or as required |
Check screen-film contact |
Semi-annually |
Or as required |
Check film speed |
Annually |
|
Check safelight |
Semi-annually |
Check safe handling time |
Check darkroom |
Semi-annually |
Check for light leaks |