Setup;

Power ON;
  • Insert the interconnect cable connector from the Workstation to the C-Arm receptacle by aligning the red dots,
  • Connect the footswitch and/or handswitch on the interface panel located on the left panel,
  • Make certain that all connectosr are locked into place
  • Connect the Workstation AC power cord to a power source dedicated to powering the fluoroscope,
  • Check that the Battery Charger indicator on the Mobile C-Arm Interface panel is illuminated ,
  • .Insert OEC key and turn to;
    • ON (clockwise position) to enable x-ray and vertical column drive,
    • Standby (counter-clockwise position) to disable x-ray and vertical column movement,
  • Press the power switch at the Workstation (located beneath the right hand monitor),
  • The OEC system begins its power up sequence,
  • Do not disturb the connectors or the Workstation keyboard while power-up is underway,
  • Power-up is complete when the Patient Information Screen is displayed on the right hand monitor, and Auto Fluoro technique appears on the C-Arm control panel,