kVp |
Half-Value Layer |
Size of X-ray Field at Entrance Surface to Patient |
|||
5 x 5 |
10 x 10 |
15 x 15 |
30 x 30 |
||
60 |
1.1 |
0.88 |
0.85 |
0.83 |
0.82 |
1.5 |
0.86 |
0.83 |
0.81 |
0.79 |
|
1.9 |
0.85 |
0.80 |
0.79 |
0.78 |
|
2.3 |
0.83 |
0.79 |
0.78 |
0.76 |
|
75 |
1.4 |
0.86 |
0.82 |
0.81 |
0.81 |
1.8 |
0.83 |
0.79 |
0.79 |
0.78 |
|
2.2 |
0.82 |
0.78 |
0.76 |
0.76 |
|
2.7 |
0.81 |
0.76 |
0.75 |
0.74 |
|
90 |
1.7 |
0.85 |
0.80 |
0.80 |
0.79 |
2.2 |
0.84 |
0.78 |
0.76 |
0.75 |
|
2.6 |
0.83 |
0.76 |
0.74 |
0.73 |
|
3.3 |
0.81 |
0.75 |
0.72 |
0.71 |
|
100 |
2.0 |
0.83 |
0.79 |
0.76 |
0.76 |
2.6 |
0.83 |
0.77 |
0.75 |
0.73 |
|
3.1 |
0.81 |
0.75 |
0.72 |
0.70 |
|
3.8 |
0.80 |
0.74 |
0.70 |
0.68 |
|
Higher kVp |
0.80 |
0.74 |
0.70 |
0.68 |
Chest Exposures
1. Position the phantom in front of and touching the chest stand and ensure that the phantom is centred to the photocells.
2. Attach the dosemeter to the arm of the phantom.
3. Set the technique normally used for a 23 cm PA Chest projection.
4. Make an exposure and record the output. Repeat twice to verify the result and reproducibility.
5. Using the inverse square law, calculate the actual PEE at 23 cm in from of the face of the chest stand.
Response to mA Change
Procedure
1. If it is possible to change the milliamperage selected, choose a mA station lower than the one which is used routinely.
2. Make an exposure and record the output and measured time.
3. If possible, select a mA station higher than the one which is used routinely.
4. Make an exposure and record the output and measured time.
5. Return to the original mA station.
Evaluate as follows:
The output should remain constant. With all other factors constant, the mAs for a given kilovoltage should not change for the same size phantom or patient. For example, doubling the milliamperagc should result in approximately the same output, but at half of the measured time. A +20 per cent tolerance is recommended. If using film, ensure that the optical density remains the same on all films.
Response to kVp Change
Procedure
1. Select a kilovoltage at least 10 kVp lower than the original setting. In mammo, a kVp change of 2 is sufficient.
2. Make an exposure and record the output and measured time.
3. Select a kilovoltage at least 10 kVp higher than the original setting. In mammo, a kVp change of 2 is sufficient.
4. Make an exposure and record the output and measured time.
5. Return the kVp to its original setting.
Evaluate its follows:
As kilovoltage increases, the output (and time) should decrease when all other factors remain the same. An important consideration, when increasing kilovoltage for this test is the minimum response time of the x-ray machine. If the output and time increase (with the increased kilovoltage), the minimum response time has probably been reached. To verify whether or not this is the case, perform this test at a lower kilovoltage. Conversely, as kilovoltage decreases, the output and time will increase. If using film, the density should remain the same.
Response to Density Change
Procedure
The density selector allows operators to control the overall density on the film. Operators should know by what percentage the overall density on the film would be changed at each density selection available.
1. Leaving all other technique factors constant, pick a density selection one step higher than the original density selection.
2. Make an exposure and record the output and measured time.
3. Pick a density selection one step lower than the original density selection.
4. Make an exposure and record the output and measured time.
5. Return the density selector to its original position.
Evaluate as follows:
As the density setting is changed, the radiation output (and time) should change accordingly (i.e., as higher densities are selected, the radiation output and density should increase) and vice versa. The exposure should change by a consistent value. Commonly, this value is about 20 per cent for each selection.
If the measured exposure does not change by a consistent value, the density selector may be malfunctioning and must be repaired.
Each facility should identify the changes expected for density selection when the unit is installed and establish specific acceptance limits.
Response to Change in Patient Thickness
Procedure
1. Select the technique factors as specified for the reproducibility or PEE measurements.
2. Increase phantom thickness by placing a copper sheet (or other material) between the phantom and dosemeter. (Use aluminum only when testing mammography).
3. Make an exposure and record the output and time measured.
Evaluate as follows:
There should be an increase in both output and time. If using film, the density should remain the same.
Back-up System
Standard
Section 13 (1) of the HARP Regulation states:
"(1)
Every diagnostic x-ray machine that is equipped with an automatic exposure control shall be equipped with,
(a) an indicator that shows when the automatic exposure control mode of operation has been selected,'
(b) a means of terminating the exposure,
(i) of an x-ray tube with a potential of less than fifty kilovolts peak, when the product of the x-ray tube current and the exposure time is 2000 milliampere-seconds per exposure, or
(ii) of an x-ray tube with a potential of fifty kilovolts peak or more, when,
(A) the product of the x-ray tube current and the exposure time is 600 milliampere-seconds, or
(B) the product of the peak x-ray tube potential, current and exposure time is sixty kilowatt-seconds per exposure; and,
(c) an indicator that warns the operator that a condition set out in subclause (b)(i) or (ii) has been reached."
Procedure
1. Select a low kVp technique.
2. Place the copper sheets and lead between dosemeter and the phantom.
3. If the back-up is set manually, select a time lower than the original time measured during the reproducibility or PEE tests.
4. Make an exposure and record the output and measured time. If the unit does not go to backup, add additional lead. If it is necessary to repeat the test, allow the tube to cool for 5 minutes. Document if back-up has been reached.
5. In a manual system the back-up may be determined by multiplying the mA used by the facility by the back-up time selected (or by calculating the kilowatt-seconds).
Note:
With manual back-up systems there must be a label indicating the maximum backup time to be used with each mA selection normally used. With manual systems, it is important to note that the time selected is not an indicator that back-up has been reached. There must be a separate indicator (audible, visible or machine shut-off) that back-up has been reached.
Evaluate as follows:
If the unit operates at a potential less than 50 kVp, calculate the mAs by multiplying the mA used by the time measured. It must not exceed 2000 mAs.
If the unit has a potential of 50 kVp or more calculate:
(i) the mAs by multiplying the mA used by the time measured at back-up. It must not exceed 600 mAs, or
(ii) back-up limited to 60 kilowatts seconds. Using the following formula (for single phase machines) the kilowatt-seconds can be calculated.
(5) kilowatt-seconds kw = (a*v)/1000
where:
kw is kilowatts
a is milliamperes divided by 1000 (or tube current in Amperes)
v is voltage and is equal to kilovoltage x 1000
(or tube voltage in volts)
To obtain kilowatt-seconds multiply the kilowatts by the time measured in seconds at back-up. For three phase units multiply this product by 0.785.
Note: This formula is particularly useful for falling-load generators.
Acceptable back-up indicators may include: a light indicator, an audible indicator or machine shut down.
Photo-Cell Variation
When changing from one cell to another or from one combination of cells to another, make exposures using the same technique factors and phantom, the radiation output should be the same for each cell(s) selection. It is recommended that each facility discuss the amount of variation programmed intemally with their service company.
Procedure
1. Test each photo-cell independently (if possible).
2. Make an exposure and record the output and measured time.
3. Change the cell selection and repeat step 2.
4. Select the third cell and repeat step 2.
5. Return the cell selection to its original position.
Evaluate as follows:
The exposure should remain the same among all cells.
Note: If a variation in output is measured between cells, discuss with the facility to confirm if the equipment has been purposefully set up that way. This is more common with dedicated chest units.