Maintenance:
Radiation
Radiation is produced in these tests. Operate the equipment in a protected area, and wear a lead apron and thyroid collar during exposure.
Do the following;
Radiation
Radiation is produced in these tests. Operate the equipment in a protected area, and wear a lead apron and thyroid collar during exposure.
Do the following;
- Position the tube head above the image intensifier,
- Position a phantom on the image intensifier (a phantom with the capability of showing high and low contrast resolution, and general imaging performance is preferable),
- Select AUTO fluoro on the control panel,
- Press the X-ray ON switch on top of the Control Panel,
- Verify that an image appears on the left monitor on the Workstation, and that it stays on-screen after fluoro is stopped,
- Connect the footswitch and handswitch,
- While pressing the exposure switch on one of these controls, verify the operation of;
- Image orientation controls (image rotate, image horizontal and vertical reversal),
- Mag mode selection (NORM, MAG1 and MAG2),
- Collimator controls (iris open/close, leaf/leaves open/close/rotate),
- Mode selection (if unit is fitted with the Vascular option),
- If the test phantom has low and high contrast test capability, check the fluoro image to determine any gross changes in image quality,
- Check the image for interference, distortion or artifacts that might detract from the diagnostic use of the device,
- Check the displayed kV and mA and compare it to a reference value set by Imaging Service to determine if any gross changes in dose rate (as set by kV and mA) have occurred,
- Document all anomalies,
- Contact a qualified service person immediately if there is loss of function or loss of image quality,