BackNext
Section 9: Computerized Tomography

THE HARP ACT EXEMPTS CT DEVICES FROM MANY STANDARDS. IN 2006, THE PROVINCIAL GOVERNMENT ASSESSED THE REGULATORY NEEDS FOR OPERATION OF CT IN ITS DIAGNOSTIC IMAGING SAFETY COMMITTEE. THAT COMMITTEE MADE A REPORT BUT AS OF OCTOBER 2010 - NO STANDARDS OR SPECIFIC REGULATIONS PERTAINING TO CT ARE ON THE BOOKS OTHER THAN THE DESIGNATION REQUIREMENT. THE FOLLOWING IS WHAT XRIS WOULD HAVE RECOMMENDED AS A STANDARD IF ANYONE HAD THE FORESIGHT TO SEE THEY WERE CLOSE TO 20 YEARS AHEAD OF THE REST OF THE WORLD.

Introduction

The X-ray Inspection Service does not specifically test computerized tomography units nor are any calibration standards specified under the HARP Regulation. However, it is expected that routine quality control or preventive maintenance be performed at least every 6 months ensuring that the manufacturers standards are met. Detailed records of these test results must be kept on site.

Inspection Procedure

In the course of an inspection the following information will be verified:
  1. The installation matches plans approved by the XRIS.

  2. X-ray operators are qualified and currently registered with the College of Medical Radiation Technologists of Ontario.

  3. Written safety procedures are in place.

  4. An estimate of potential exposure to personnel (estimate may be obtained by taking scatter readings usually in the control booth and at the foot end of the x-ray table).

  5. Visual Indicators: warning sign(s), Line "ON", X-ray "ON", kVp and mA selectors. All controls must be clearly identified, emergency "OFF" button.

  6. Ensure that quality assurance program is in place and that records are kept. The records should include the results of the tests listed below.

    Every six months, or more often depending on manufacturer's instructions and standards, check for:


    • a. Sensitivity Profile
    • b. kVp Accuracy
    • c. Spatial Resolution
    • d. Plane Indication and Alignment
    • e. Scan Increment Accuracy
    • f. Laser Alignment
    • g. Thickness of Cut
    • h. Position Indicator Accuracy
    • i. Scaling (Distance) Accuracy
    • j. Noise/Resolution


    Maintain records on file which document:
    • a. The CTDI Index
    • b. Log Book for Repairs
    • c. Procedural Manual. This manual should include the following:
      • List the scan protocols to be used
      • Policies on the use of oral contrast
      • Use of intravenous contrast
      • Slice thickness, slice spacing, anatomical description of the extent of the field for the scanogram
      • The technical factors selected for the examination
      • Patient and staff shielding policy
      • Restraining policy
      • Pregnancy policy


  7. Darkroom standards (if applicable).

  8. Condition of patient shielding.
 

Acceptance Tests Requirements

For any particular unit, it is important to establish the level of performance during acceptance testing of the unit. An appropriate quality control program will ensure maintenance of this performance level over time. At acceptance testing, the manufacturer is required to furnish specific dose information obtained by using the CT dosimetry phantom and imaging performance information.

Following the installation of new or used CT scanners, acceptance tests must be performed and submitted to the X-Ray Inspection Service within sixty days of the installation. Acceptance tests should include, but not necessarily be limited to, the following information:
  1. Installation conformity to approved plans for radiation safety.
  2. Actual Patient Entrance Exposure (Patient Entrance Exposure) measurements for examinations most commonly performed (i.e., head and abdomen):
    1. These measurements should be compared to the dose index profile (CTDI) provided by the manufacturer.
    2. Results should include measurements taken at 3 different MAS values for 3 different slice thickness available (if applicable).
  3. Note:

    The RPO of the facility should decide on how these patient dose measurements are to be done as long as the test procedure can be duplicated at a later date to verify consistency in output.
  4. Results for all the tests listed under Inspection Procedure in item 6.
 

Additional Tests for Computerized Tomography

Table